Troubleshooting, Storage and shipping – BNC 1621M User Manual

Page 22

Advertising
background image

4. TROUBLESHOOTING

The list of possible problems and their solutions are specifi ed in Table 4.1

5. STORAGE AND SHIPPING

5.1. Packaged devices may be shipped by any method of closed transport

at the air temperature from -50 to +50°C .

5.1. Packaged devices may be shipped by any method of closed transport

5.1. Packaged devices may be shipped by any method of closed transport

5.2. Packaged devices should be secured during transport.

5.3. When carried by sea, packaged devices should be sealed in

plastic bags with silicagel.

5.4. When carried by air, packaged devices should be placed

in sealed compartments.

5.5. Devices should be stored in the manufacturer’s package at an air

temperature ranging from -15 to +50°C and relative

5.5. Devices should be stored in the manufacturer’s package at an air

5.5. Devices should be stored in the manufacturer’s package at an air

humidity up to 95% at a temperature of 35°C.

temperature ranging from -15 to +50°C and relative

temperature ranging from -15 to +50°C and relative

5.6. Devices without package should be stored at an air temperature from

10 to 35°C and relative humidity up to 80% at a temperature of 25°C.

5.6. Devices without package should be stored at an air temperature from

5.6. Devices without package should be stored at an air temperature from

5.7. The storage location should be free of dust, chemical vapors,

aggressive gases and other substances that may cause corrosion.

Note: If a defective battery is used the Er1-Er7 message may appear.

Attention! When the Er1-Er7 message appears, press any button. When the error message appears for the

A

A

second time (approx. 15 min) the dosimeter is not to be used.

Problem

The LCD indicates “bAt”

message

No indictions on LCD

The device does not respond

to pressing a button. the LCD

indicates incorrect symbols

LCD indicates Er1-Er7

Possible Cause

Battery discharge

Battery discharge

Battery inserted incorrectly

Microprocessor error condition

Dosimeter failure

Solution

Replace Battery

Replace Battery

Insert battery correctly

Remove battery and

insert again after 5 minutes

Return instrument for repair.

21