Installation - site requirements, Jair3 and jair5 jair7 – JUN-AIR JAir Dental Compressed Air System User Manual

Page 10

Advertising
background image

10

CI-101A (Rev B)

JUN-AIR Dental Compressed Air System User Guide

© 2011, JUN-AIR

We reserve the right to make any alterations which may be due to any technical improvements

Printed in the USA

INSTALLATION - SITE REQUIREMENTS

Site Requirements

JAir3

JAir5

JAir7

Electrical
Voltage min./max*

208/253

208/253

208/253

Full-load amps

7 .5

11

15 .0

Minimum circuit breaker rating amps**

20

20

30

Min . wire size (AWG)

12

12

10

* Install a buck-boost transformer if electrical service is below 208 volts or above 253 volts.
Contact authorized technical service for recommended buck-boost transformers.
** Circuit breaker must isolate all poles simultaneously.

Minimum of 12-inch clearance

between unit and wall - all around

Minimum of 12-inch clearance

between unit and ceiling

19.8

28.9

15.2

17.6

Hose barb for drain

(accepts 1/4 I.D. tube)

16.0

28.0

27.8

System exhaust to operatory

(3/8 female NPT) utilize 1/2-inch

copper piping for operatory.

Flexible connection hose available to

connect system with operatory copper piping

Minimum

of 12-inch

clearance

between

unit and

wall

Minimum

of 12-inch

clearance

between

unit and

wall

Minimum

of 12-inch

clearance

between

unit and

wall

Minimum of 12-inch

clearance between unit

and ceiling

Remote leads - 18 AWG

(red, yellow, blue)

Main power leads

(black, black,

green/yellow)

Intake filters

11.8

JAir3 and JAir5

JAir7

Minimum of 12-inch clearance

between unit and wall - all around

Minimum of 12-inch clearance

between unit and ceiling

Minimum

of 12-inch

clearance

between

unit and

wall

Minimum of 12-inch clearance

between unit and ceiling

Minimum

of 12-inch

clearance

between

unit and

wall

Minimum

of 12-inch

clearance

between

unit and

wall

Remote leads - 18 AWG

(red, yellow, blue)

Main power leads

(black, black,

green/yellow)

Intake filters

12.5

Hose barb for drain

(accepts 1/4 I.D. tube)

34.1

System exhaust to operatory

(3/8 female NPT) utilize 1/2-inch

copper piping for operatory.

Flexible connection hose available to

connect system with operatory copper piping

18.0

37.0

27.3

35.7

19.0

20.5

23.7