IKA C 1 Package 1/10 User Manual

Page 20

Advertising
background image

20

Prerequisite checking for a measurement

When checking the preconditions for a measurement using a

distinction is made with and without C1.20 heating between. The

following points should be checked:

• Existing Ignition

• Sufficient oxygen present

• Open water connection

• Tempered water or using a heating their functionality

• Ignition:

Checking whether ignition contact available

• Oxygen pressure:

Check whether the system has sufficiently high oxygen pressure

has to make a measurement

• Water detected:

Check whether the system has detected water

• Limit ≤ Water ≤ Start temp:

Check whether there is the current water temperature within the

permissible temperature range

Without

C1.20

heating:

To increase the service life of wearing parts (O-rings, seals,

etc.) it is advisable to always work with a water template.

Generally, the test sample must be selected so that the tempe-

rature elevation during the measurement is below 10 K and

the temperature increase of the calibration comes close (max.

energy input: 40.000 J). Such exposure may damage the

calorimeter.

When exceeding the maximum energy input is recommended

to return the calorimeter.

To determine the potential energy initially very small sample

weights (approx. 0.2g) must be chosen when working with

unknown substances. If you burn unknown samples, leave

the room or keep your distance from the calorimeter.

If distilled water or solutions presented in the decomposition

vessel during the combustion experiment, the calibration

with the same template must have been performed before.

• Heater test:

Check the heating functionality

With

C1.20

heating:

Close the calorimeter

After the sample is loaded and the contact with the ignition wire

is present and the inner vessel was placed, the calorimeter is

sealed.

The following screen informs the user whether the system is

closed properly and securely.

Are combustion vessel and the complete system properly closed,

the measurement starts automatically.

C1 092014

Note:

Can the inner container and / or closure ring with handle heavy

wire, please follow the instructions for oiling the seals! (Page 74)