Measurement procedure

During a measurement, different phases are passed through.

These phases are described below

Phase 1:

Filling the calorimeter with water

Phase 2:

Filling the calorimeter with oxygen (Note: If the option-O

2

flushing under the main menu " Basic Settings is selec-

ted, the inner vessel is two times with Oxygen rinsed)

NOTE:

Alternatively, to display the graph may also have a detailed

view of be selected with the following information:

Current temperature:

Mean value of the actual temperatures

Ignition temperature:

the currently measured temperature

Measurement time:

Aisplay: duration of the measurement

NOTE:

After switching on the calorimeter C1 or after a break of 2

hours during the filling of the calorimeter C1 is still carried

out an intermediate emptying (H

2

O - purge). Once the

water sensor detects water, the stirrer is switched on.

After a short waiting period begins phase 2.

C1 092014

Advertising