3 test instructions, lower fluorescence, Test instructions, lower fluorescence, Mean 100 * stdev (cv%) precision – Eppendorf AF2200 Plate Reader User Manual

Page 55

Advertising
background image

Maintenance

Eppendorf

®

PlateReader AF2200

English (EN)

55

Calculating the precision from one measurement to the next

The result of the formula specifies the precision from one measurement to the next in % CV (coefficient of
variation).

8.6.3

Test instructions, lower fluorescence

The following tests can be conducted on the device to check the device specifications for this measuring
procedure:

• Sensitivity

• Precision within a plate

• Precision from one measurement to the next

Sensitivity

Conduct the measurement with the following settings to determine the detection limit for fluorescein.

Mean value

Well A1

Mean RFU value of well A1 over 20 kinetic cycles

Std. deviation

Well A1

Standard deviation of RFU values from well A1 over
the 20 cycles

Parameter

Setting

Measuring mode

Lower fluorescence

Wavelength, excitation

485 (20 nm)

Wavelength, emission

535 (25 nm)

Number of flashes

25

Integration time

40 (μs)

Setting time

0 (μs)

Signal gain

Optimal

Plate type

GRE96fb

wellA1

wellA1

mean

100

*

stdev

(CV%)

Precision

=