7 absorbance, 8 "quick" measurements, Absorbance – Eppendorf AF2200 Plate Reader User Manual

Page 64: Quick" measurements

Advertising
background image

Technical data
Eppendorf

®

PlateReader AF2200

English (EN)

64

9.7

Absorbance

The following specifications apply for a

• wavelength range of 300 – 700 nm and

• measurements that are conducted with 25 flashes (measurements) per well.

9.8

"Quick" measurements

"Quick" measurements are the fastest measurements that can be performed with the device. They are
performed using only one flash.

• 96-well plates (FI, absorbance) measuring time: 20 s

• 384-well plates (FI, absorbance) measuring time: 30 s

Parameters

Characteristics

Wavelength range

230 - 1000 nm

Measuring range

0 – 4 OD

Plate type (number of wells)

96

Accuracy 0 – 2 OD

< ± 1% + 10 mOD

Accuracy 2 - 3 OD

< ± 2.5%

Basic line stability

± 10 mOD (1 Sigma)

Wavelength precision

Depends on filters used

The time for moving the plates in and out is not taken into account when specifying the
measuring time.