7 absorbance, 8 "quick" measurements, Absorbance – Eppendorf AF2200 Plate Reader User Manual
Page 64: Quick" measurements
Technical data
Eppendorf
®
PlateReader AF2200
English (EN)
64
9.7
Absorbance
The following specifications apply for a
• wavelength range of 300 – 700 nm and
• measurements that are conducted with 25 flashes (measurements) per well.
9.8
"Quick" measurements
"Quick" measurements are the fastest measurements that can be performed with the device. They are
performed using only one flash.
• 96-well plates (FI, absorbance) measuring time: 20 s
• 384-well plates (FI, absorbance) measuring time: 30 s
Parameters
Characteristics
Wavelength range
230 - 1000 nm
Measuring range
0 – 4 OD
Plate type (number of wells)
96
Accuracy 0 – 2 OD
< ± 1% + 10 mOD
Accuracy 2 - 3 OD
< ± 2.5%
Basic line stability
± 10 mOD (1 Sigma)
Wavelength precision
Depends on filters used
The time for moving the plates in and out is not taken into account when specifying the
measuring time.