5 troubleshooting, 1 error search, Eppendorf® uvette® — instructions for use – Eppendorf UVette User Manual

Page 8

Advertising
background image

8

Eppendorf® UVette® — Instructions for use

5

Troubleshooting

5 Troubleshooting

5.1 Error search

Symptom/

message

Cause

Remedy

Measured values

are imprecise.

The blank value and sample

value have been measured

in different cuvettes.

Measure the blank value and

sample value in the same

UVette.

The position of the UVette in

the device is not correct.

UVette was inserted

incorrectly.

Press the UVette firmly onto

the base of the cuvette shaft

or adapter.

The UVette must be inserted

in the device in the same

direction for the blank

measurement and sample

measurement. The UVette

may not be rotated 180°.

The UVette was used for

more than one

measurement.

Each UVette may only be

used for one measurement

(blank measurement or

sample measurement).

Measured values

are not plausible.

The light path height of the

UVette does not match the

light path height of the

device.

Use the adapter (not with the

Eppendorf BioPhotometer

or Eppendorf

BioSpectrometer).

The adapter may only be

used with the light path

height that matches the

photometer.

Optical path length of the

UVette does not match the

value set in the device.

Insert the UVette in the

device so the required

optical path length (2 mm/10

 mm) is penetrated.

If the optical path length has

been changed, the blank

value also needs to be

updated.

Set the optical path length

parameter in the device

(2 mm/10  mm).

The wavelength is below

220 nm or above 1 600 nm.

The UVette may only be

used for wavelengths of

220 nm to 1 600 nm.