5 operation, Operation – Eppendorf Temperature Verification System User Manual
Page 19
Temperature Verification System USB – Single-channel — Operating manual
19
5
Operation
5 Operation
CAUTION!
CAUTION! Risk of burns on thermoblock, heated lid and temperature
sensor.
The thermoblock, heated lid and temperature sensor can reach temperatures
of more than 95 °C.
During gradient operation, the temperature of the heated lid and thermoblock
is higher on the right side than on the left.
Use extreme caution when conducting measurements on the heated
thermoblock.
WARNING!
WARNING! Burns from hot surfaces.
The metal surfaces of the exchangeable thermoblock and the heating plate
reach temperatures up to 100 °C.
Do not touch hot surfaces.
Wait until the thermomixer or thermostat have cooled to room
temperature.
Test, service or repairs may only be conducted after the devices have
cooled to room temperature.
NOTICE!
NOTICE! Damage to the sensor/cable connection as a result of the
cable being pulled.
Repeatedly pulling on the cable to remove the temperature sensor from the
thermoblock will cause damage to the sensor/cable connection.
Pull the thermosensor out of the thermoblock via the thermosensor
housing.
Do not apply tractive forces to the cable.