If there is doubt that dispensing data are correct, 7 troubleshooting, Troubleshooting – Eppendorf Research pro User Manual

Page 32

Advertising
background image

32

If there is doubt that dispensing data are correct

To avoid dispensing errors, the precision and accuracy of the Research pro need to be checked
regularly. The PICASO II software program (see ordering information) is available to determine
permitted systematic and random measuring deviation (see Section 2 "Technical data"). An
SOP (

S

tandard

O

perating

P

rocedure) for checking pipettes can be called up from our home

page

www.eppendorf.com

.

For liquids whose physical data deviate significantly from those of water, calibration needs to be
changed in accordance with Section 5.11.1. Calibration will also need to be changed if the
pipette is being used at a very high altitude.

Program notes only for program version prior to V. 1.56

Error

Cause

Solution

(Continued)
Multi-channel:
Pipette doesn’t
aspirate liquid.

Push the piston actuator back
into the upper position using
side "B"of the universal tool
and then remount the ejector.

Following assembly:

Trigger a motor reset by
holding down the Reset key
(until

appears in the

display and the motor
moves).

Error

Cause

Solution

The program level cannot
be opened.

Programm version of the pipette
does not match the description.

With the old program version,
the program level was opened
by pressing the key once.

The ASP program cannot
be found.

Programm version of the pipette
does not match the description.

With the old program version,
the ASP program did not
exist.

VOL was not found among
device parameters.

Programm version of the pipette
does not match the description.

With the old program version,
the parameter was called
OPT; A choice between two
volumes can be made by
using Select.

R

7 Troubleshooting

Troubleshooting

7

E06 Pflege etc. Seite 32 Dienstag, 19. Dezember 2006 2:16 14

Advertising