1 safety precautions and applicational limitations, Intended use, Handling – Eppendorf Research pro User Manual

Page 4: Care and maintenance, Battery, Safety precautions and applicational limitations

Advertising
background image

4

Intended use

The Research pro is a lab device intended for dispensing liquids in the volume range from
0.5 µL to 5 mL, in combination with matching pipette tips.
In vivo applications (applications in or on the human body) are not permitted.
The Research pro may only be operated by trained specialist staff. All users must have
read the operating manual carefully and familiarized themselves with the device's mode of
operation.

To guarantee problem-free, safe operation of the pipette, it is essential to observe the following
points:

Handling

Use the Research pro only when a pipette tip is attached.

Pipette tips are solely designed for disposable usage.

Do not lay down the Research pro when a filled pipette tip is attached.

After liquid has been aspirated, press the Actuate key only when you are sure that the liquid
shall be dispensed correctly.

When the pipette is switched on and when the Reset key is pressed, the piston may move.
Please ensure beforehand that the liquid in the pipette tip shall be dispensed correctly.

Emergency stop: A moving piston can be stopped by pressing the Reset key.

When using infectious, radioactive, toxic and other solutions which may pose health risks,
please observe the safety precautions laid down for your country.

Do not use the Research pro in a potentially explosive environment or with potentially
explosive chemicals.

When using organic solvents and aggressive chemicals or their vapors, please check their
compatibility with the pipette tip (PP) and pipette.

When using solutions with physical data which deviates greatly from those of water, carry
out the procedure for checking the dispensing volume, as described in Sec. 5.11.1.

Operate the Research pro at temperatures between 15 °C and 40 °C only and at a humidity
of max. 80 % at a temperature up to 31 °C.

When using Filtertips or 200 µL tips with the 300 µL pipette, observe the volume restriction
(see Sec. 5.11).

Care and maintenance

Do not clean the display or any of the labeling using acetone or aggressive solutions.

Do not allow any liquid to enter the device.

Repairs may be carried out by authorized service personnel only.

Use original spare parts and accessories (battery, pipette tips) only.

Battery

Charge up the battery before operating the device for the first time (see Sec 3).

Charge the battery in the pipette only, using the charging adapter or the charging stand and
the original power unit.

If a flashing battery symbol appears in the display, stop dispensing and charge the battery.

When changing the battery, do not allow the battery to come into contact with any metallic
surfaces.

Dispose of used batteries as special waste, in accordance with legal stipulations to this
effect. Batteries may not be disposed of together with household waste!

If rechargeable batteries are damaged, avoid touching them with the hands.

Never charge the Research pro without a battery.

Before storing the Research pro for a long period, be sure to remove the charged battery.

When the Research pro is stored for a long time, charge the batteries approximately every
six months.

Safety precautions and applicational limitations

1 Safety precautions and applicational limitations

1

E01 bis 04 Seite 4 Dienstag, 9. Februar 2010 5:49 17