Innova® -86 °C Freezers — Operating manual

6.4

Alarm monitoring socket

The freezers are provided with an alarm monitoring socket at the rear of the freezer and a
matching plug for external monitoring. This plug can be connected either to a central monitoring
system such as a building management system, or to a remote alarm via an auto-dialer.

Abb. 7:

Upright freezers alarm monitoring socket (except U101)

Fig. 7:

Upright freezers alarm monitoring socket (except U101)

Abb. 8:

Chest (and U101) freezers alarm monitoring socket

Fig. 8:

Chest (and U101) freezers alarm monitoring socket

The configuration of the socket is shown in (Fig. 9 on p. 26) and (Fig. 10 on p. 26), as viewed
from the rear of the freezer. Within the freezer, the socket is connected to voltage-free contacts
rated at 24 volts, 1 amp. In normal operation, with the mains/power on, pin 1 is connected to pin
2 (N/C), and in the alarm condition, with mains/power off, pin 1 is connected to pin 3.

The High Temperature Alarm output to the Remote Alarm Monitoring Socket can be programmed
to a set time delay (see Setting the alarm delay on p. 23).

1

RS-485 connector (optional)

2

Alarm monitoring socket

3

Mains/power socket

1

RS-485 connector (optional)

2

Alarm monitoring socket

3

Mains/power socket

1

2

3

1

2

3

NOTICE!

Risk of material damage



Hazardous voltages must not be connected to the remote alarm socket. Max Rating 24 V 1 A.