Technical data

9

Technical data

9.1

Specifications

9.1.1

Upright freezers specifications

* X = 0 for 120 V, 2 for 208 - 230 V (N/A for U101), or 1 for 230 V

‡ Freezer set to -80 °C, ambient 20 - 25 °C at rated mains/electrical supply

Model No.

U101

U535

U725

Part No.

U9420-000X*

U9430-000X*

U9440-000X*

Internal Dimensions:

Height x Width x Depth

640 x 480 x 330 mm

25.2 x 18.9 x 13 in

1365 x 640 x 615 mm

53.7 x 25.2 x 24.2 in

1365 x 865 x 615 mm

53.7 x 34.0 x 24.2 in

External Dimensions:

Height x Width x Depth

830 x 900 x 566 mm

32.7 x 35.4 x 22.3 in

1950 x 800 x 867 mm

76.8 x 31.5 x 34.1 in

1950 x 1025 x 867 mm

76.8 x 40.4 x 34.1 in

Capacity

101 Liters

3.6 cubic feet

535 Liters

18.9 cubic feet

725 Liters

25.6 cubic feet

Net Weight

116 kg

256 lb

250 kg

551 lb

315 kg

693 lb

Lock

Standard

Standard

Standard

No. Compartments

2

3

3

Interior

Stainless steel grade 304L

Alarms

High/Low temperature, power fail, battery low, filter clean, fault

Insulation Material

Vacuum insulation panels and urethane foam

Remote alarm port

Standard

Standard

Standard

RS-485 interface

Optional

Optional

Optional

Refrigerants:

High Stage Refrigerant: R404A / Low Stage Refrigerant: R508B

‡Power Consumption:

120 V electrical supply

265 Watts

550 Watts

N/A

208 - 230 V electrical
supply

N/A

550 Watts

685 Watts

230 V electrical supply

367 Watts

550 Watts

660 Watts

Mains/Power Source and Current Rating:

120 V, 60 Hz

13 A

16.5 A

N/A

208 - 230 V, 60 Hz

N/A

9 A

10 A

230 V, 50 Hz

5 A

5 A

9.5 A

Pull Down Time: From +25 °C to -85 °C (freezer empty; 240 V, 50 Hz electrical supply)

3.7 hours

5.3 hours

5.6 hours

Performance

-50 °C to -86 °C at 32 °C maximum ambient operating temperature

Environmental Conditions All freezers use components tested to CE/UL specifications listed below:

Indoor use

Altitude up to 2000 m

Ambient temperature range 10 °C to 32 °C

Maximum relative humidity 80 % for temperatures up to 31 °C, decreasing linearly to
50 % relative humidity at 40 °C

Mains/power supply voltage fluctuations not to exceed ± 10 % of the nominal voltage

Installation category II

Pollution degree 2

8