4 testing back-up system operation, 5 battery back-up, Testing back-up system operation – Eppendorf CO2 Back-up System User Manual

Page 22: Battery back-up

Advertising
background image

Operation
New Brunswick CO2 Back-up System
English (EN)

22

2. Check for frozen condensation around injector and freezer shelf. Frozen condensation indicates CO

2

Back-up System is working.

3. Push

Valve Test button.

Gas should

not be injected into freezer.

Valve Open light should not illuminate.

7.4

Testing back-up system operation

1. Turn off CO

2

Back-up System with the

System On switch.

2. Turn on freezer.

3. Set freezer to to desired temperature and wait until temperature is reached.

4. Set

Temp Set dial on the control unit to 10 °C higher than the freezer's temperature.

5. Flip

System On switch on.

The

Temperature Warning LED and Valve Open LED remain off. This is because the temperature of

freezer is lower than temperature set on

Temp Set dial.

6. Change

Temp Set dial to 10 °C lower than freezer's temperature .

The

Temperature Warning LED will turn on indicating that the temperature in the freezer is below the

temperature set on the dial of the control unit.
The

Valve Open LED will flash on and off indicating gas is being injected into the freezer.

7. Unplug CO

2

Back-up System from mains.

The

Power On LED will turn off. The device is now operating on battery back-up. There should be no

change in temperature in the freezer while CO

2

Back-up System is in operation.

8. Plug the CO

2

Back-up System back into mains.

9. Reset the

Temp Set dial to the desired temperature at which the device should turn on in the event of

power failure.

10. Verify that the gas supply is still on and the freezer is operating.

The

Temp Warning should be off indicating that the freezer temperature is below the temperature set

on the

Temp Set knob.

The

Valve Open LED should be off indicating that gas is not being injected into the freezer.

The

Power On LED should be on indicating that the CO

2

Back-up System is plugged into the mains.

7.5

Battery back-up

The CO

2

Back-up System contains a rechargeable battery in case of power failure.

If the CO

2

Back-up System is plugged into the mains, and the

Power On LED is illuminated, the battery is

being charged.

If the external power supply fails, the back-up battery will power the CO

2

Back-up System for up to 48

hours, depending upon ambient conditions. The lower the ambient temperature, the longer the back-up
battery will run.