6 alarm system monitoring socket, Alarm system monitoring socket – Eppendorf CO2 Back-up System User Manual

Page 23

Advertising
background image

Operation

New Brunswick CO2 Back-up System

English (EN)

23

When the back-up battery's voltage drops below 11.4 volts, the

Battery Low LED will illuminate and an

alarm will sound.

7.6

Alarm system monitoring socket

The CO

2

Back-up System is designed with a fixed socket at the rear labelled

Alarm Socket, and a plug for

external monitoring purposes.

Within the control unit, the alarm socket is connected to volt-free, switching contacts rated at 24 volts, 1
amp. In normal operation, pin 1

(E) is connected to pin 2

(L) and in the alarm condition, pin 1

(E)

is

connected to pin 3

(N).

The alarm plug can be used to connect to a remote alarm device or central monitoring system.

WARNING! Hazardous voltages

Hazardous voltages must not be connected to the remote alarm socket. Maximum rating is
24 volts, 1 amp.

NOTICE!

External device shall secure double reinforced insulation from mains voltage in accordance
with 61010-1

1

Freezer Socket

2

Alarm Plug (external view)

3

Alarm Plug (internal view)

1

2

3