Hef® and g -86 °c freezers — operating manual, Op erating control s and function – Eppendorf New Brunswick -86 °C Freezers G User Manual

Page 18

Advertising
background image

HEF® and G -86 °C Freezers — Operating manual

18

Temp-alarm light

Power fail light

Fault light

Battery low light

Filter clean light

Remote control light

Item

Name

Function

8

TEMP-ALARM Should a mains/power failure cause the temperature to surpass

the alarm setpoint, the TEMP-ALARM illuminates. (The audible
alarm will already be sounding due to the mains/power failure).
The TEMP-ALARM lamp will extinguish when the temperature
returns to normal set range. Cancel the TEMP-ALARM by
pressing the ALARM TEST/MUTE KEY.

Item

Name

Function

3

POWER-FAIL

Illuminates if the mains/power supply fails, flashing at
approximately 10-second intervals, accompanied by an audible
alarm. When mains/power is restored, the indicator goes off
and the audible alarm stops. (The battery must be switched on
and charged for this indicator to operate.)

Item

Name

Function

4

FAULT

Illuminates if there is a system failure within the freezer.
Interfacing with the S.M.A.R.T. Plus™ diagnostics via the
control panel, the fault can be determined (see Error messages
on p. 34)
. Sy
stem failure is accompanied by an audible alarm.
Correction of the fault extinguishes the light and audible alarm.

Item

Name

Function

5

BATTERY-LOW With mains/power ON: illuminates if battery voltage is below 6

volts, flashes when voltage drops to 5 Volts. With mains/power
OFF: if battery voltage drop below 5.5 Volts, this fault indicator
will stop functioning.

Item

Name

Function

6

FILTER-CLEAN Illuminates, accompanied by an audible alarm, to indicate a

blocked or dirty filter. Filter is located on the front at the bottom
of all freezers. Remove by turning the two thumbscrews on the
filter holder ¼ turn. Clean filter by washing in mildly soapy
water, then air dry. If filter warning light does not go out after
replacing the cleaned filter, contact your local New Brunswick
service representative.

Item

Name

Function

12

REMOTE

CONTROL

Indicates when freezer is operating under remote computer
control via the optional RS-485 interface port and New
Brunswick's BioCommand

®

SFI software, or other laboratory

data logging software.

5

Op

erating

control

s and function

Advertising