Eppendorf R Series 48R CO2 Incubator User Manual

Page 20

Advertising
background image

20

CO48R-0050

Š Galaxy R Series 48 R Personal Size CO

2

Incubator

User’s Guide

NOTE:

Be certain to check for leaks in the CO

2

connections to avoid depleting

your CO

2

gas supply. This can be accomplished using a solution of

soapy water applied to each fitting and checking for bubbles. If any
bubbles are noted, readjust the fitting.

6. The chamber setpoints are pre-programmed at 37.0°C and 5% CO

2

. Leave the

incubator on until the programmed chamber temperature and CO

2

concentration has

been reached (please see NOTE below).

NOTE:

The incubator’s CO

2

valve is disabled until the incubator reaches the

temperature setpoint. After reaching the temperature setpoint, the CO

2

valve is activated, allowing the incubator to reach the CO

2

setpoint. If

power is interrupted to the incubator long enough for the temperature to
drop below setpoint, the CO

2

valve will be deactivated until setpoint is

again achieved. (This serves to avoid spurious CO

2

readings while the

incubator is reaching its temperature setpoint)

7. Leave the incubator running for at least two hours (preferably overnight) to allow

conditions to stabilize.

3.3 Using the Humidity Tray & Humidity Control

If humidification is required, the humidity tray should be filled with 0.5 liters of warm
(

∼ 37.0ºC) distilled water at this time.


For cell culture work, we recommend the use of copper sulphate (or a recognized
biocide) in the humidity tray. Tests have shown that, in addition to inhibiting bacterial
growth in the tray, this can reduce contamination on the chamber walls. Add one small
teaspoonful (~0.11 oz or 3.2 g) of copper sulphate each time you fill the tray with fresh
water.

For IVF and other sensitive work, we do not recommend the use of any biocide in the
humidity tray. To reduce the possibility of contamination, every 10 to 14 days, empty
the tray, clean it with a solution of 70% isopropyl alcohol and 30% distilled water, and
then refill it with 0.5 liters of warm distilled water.

To remove the humidity tray, turn it 45º to clear the shelf rack supports (see Figure 3a).