enter into the sterilization procedure. The Default time is as following:

1. Modo 1

Vacuum: 10 min

Ozone generating: 30 min

Ozone-maintain: 10 min

Ozone decomposition: 20 min

The whole sterilization time is 70 min.

2. Modo 2

Vacuum: 10 min

Ozone generating: 25 min

Ozone-maintain: 10 min

Ozone decomposition: 15 min

The whole sterilization time is 60 min.

NOTES:

 After start, it shows TIME and SIGN of each stage by turns. The

procedure and sign of each stage is as following:

Vacuum (CC)  Ozone generating (O3)  Ozone-maintain (OF) 

Ozone decomposition (O2)  OVER (OP).

 During the auto mode procedure, any press on thin film will be no

function.

 It shows

“OP” when whole procedure is completed and buzz 3 times

at the same time. After, turn off the main switch.