3.2 Preparation/Cleaning of Glass Plates

Hand wash both plates with a high quality lab detergent followed by a complete rinsing
with dH

2

O. Air dry or use a lint-free tissue. Spray/wipe the chosen inner surfaces of the

plate set with 95% ethanol and dry with lint-free tissue.

3.3

Gel Casting Using Gel Wrap Gasket Casting method

1. Start by holding the rectangular back plate with

the rounded bottom corners and start applying
the gasket around one side of the glass plate.
Note: one side of the “U” shaped gasket is flat,
and the other side has tubing that will act as a
seal around the spacers.




2. When applying the gasket over the rounded

corners of the back glass plate, make sure the
notches on the gasket align with the rounded
corners of the glass plate. Once the gasket is
pushed over the bottom edge and corners, work
it down the remaining side.




3. Place the gasketed plate on the lab bench with

the tubing side up, and extend the bottom of the
plate over the edge of the bench, approximately
¾ of an inch. Place the spacers along side the
inside edges of the gasket. Be sure the rounded
corner end of each spacer is facing the outside
bottom of the plate, following the radius of the
glass.











4. Place the notched plate on top of the bottom

assembly, starting from the bottom edge and
gently easing the plate down. Verify the gasket is
smooth around the edges and then clamp along
the bottom.