Large Format Vertical

Section 3
Instructions for Use

3.1

Large Format Vertical Unit Preparation

A.

Unit Set-Up
1. Place vertical unit on a level work surface in an authorized work area. Attach each of

the black guide rods to the flathead screws through the countersunk holes on the
bottom of the base. Thread the screws into the black guide rods and hand tighten.
Do not over tighten.


2. Slide the upper reservoir onto the guide rods. Be sure to unscrew the thumbscrews

slightly so that the ends of the screws are not protruding into the guide rod channels.

3. Adjust the upper reservoir so that the facia and glass notches are level, using a

notched plate as a guide. Tighten the two black thumbscrews to fix the height.

4. Level the unit by adjusting the four white nylon thumbscrews until the bubble is

centered in the level. Be sure that all four of the screws are touching the bench.

5. Verify that the comb, spacer set and Gel Wrap gasket are the same thickness by

assembling without gel casting.


3.2 Preparation/Cleaning of Glass Plates

Hand wash both plates with a high quality lab detergent followed by a complete rinsing
with dH

2

O. Air dry or use a lint-free tissue. Spray/wipe the chosen inner surfaces of the

plate set with 95% ethanol and dry with lint-free tissue.

3.3

Gel Casting Using Gel Wrap Gasket Casting method

1. Start by holding the rectangular back plate with

the rounded bottom corners and start applying
the gasket around one side of the glass plate.
Note: one side of the “U” shaped gasket is flat,
and the other side has tubing that will act as a
seal around the spacers.




2. When applying the gasket over the rounded

corners of the back glass plate, make sure the
notches on the gasket align with the rounded
corners of the glass plate. Once the gasket is
pushed over the bottom edge and corners, work
it down the remaining side.




3. Place the gasketed plate on the lab bench with

the tubing side up, and extend the bottom of the
plate over the edge of the bench, approximately
¾ of an inch. Place the spacers along side the
inside edges of the gasket. Be sure the rounded
corner end of each spacer is facing the outside
bottom of the plate, following the radius of the
glass.

Advertising