Installing Drain Water Pump (PUMP301)

In applications where a floor drain is not available and a CARON water recycling

system is not being used, a drain pump can be purchased to pump any excess

condensate from the chamber to a local sink or drain. The pump is located near the

middle of the back of the chamber. Connect the supplied tubing from the pump to the

sink / drain. The tubing may be run vertically into a ceiling but should not exceed 15

feet height.

Installing Side Mounted Recorders (RCDR320, RCDR321, RCDR322 & RCDR323)

The recorder will arrive packaged inside the chamber. Carefully remove the recorder

from its packaging. Mount the recorder by using the pre-installed recorder bracket.

There are three factory drilled holes located on the right side of chamber as you face

the front of chamber. Using the factory supplied screws, screw the recorder to the side

of the chamber.

There are two cables that come out of the recorder. One is to power the recorder; the

other is the temperature and/or humidity signals coming from the chamber. With

power to the equipment turned off, plug the two connectors into their mating connector

at the top of the chamber. Turn power to the chamber back on.

Standard factory set up for chart speed is 7 day operation. Refer to the Chart

Recorder’s User’s Manual provided with the recorder to change the chart speed

settings for various chart speeds.