7000-50 & 7000-75 Series Operations Manual Rev A 1/6/2014

Page 39 of 71

SECTION 7 – CALIBRATION

The temperature and humidity systems can all be calibrated as necessary. CARON

recommends an annual calibration check of each system. Before making a calibration

adjustment, allow the cabinet to stabilize a minimum of 12 hours from a power off

condition. If the unit has been in operation, allow a minimum of 3 hours of stable

operation at all set-points. If you do not have the appropriate reference instruments to

perform calibration, contact CARON’s service department for on-site calibration at

[email protected]

. Caron also provides validation services which ensures

that the unit is functioning properly according to IQ, OQ and PQ protocols which satisfy

FDA guidelines for qualification verification of equipment.

The Calibration Screen

To get to the calibration screen from the home page:

Press the

(Settings) button.

Be sure that all reference instruments are

calibrated to an appropriate standard.

Settings

Button