Caron 7000-75 User Manual
Page 39
7000-50 & 7000-75 Series Operations Manual Rev A 1/6/2014
Page 39 of 71
SECTION 7 – CALIBRATION
The temperature and humidity systems can all be calibrated as necessary. CARON
recommends an annual calibration check of each system. Before making a calibration
adjustment, allow the cabinet to stabilize a minimum of 12 hours from a power off
condition. If the unit has been in operation, allow a minimum of 3 hours of stable
operation at all set-points. If you do not have the appropriate reference instruments to
perform calibration, contact CARON’s service department for on-site calibration at
. Caron also provides validation services which ensures
that the unit is functioning properly according to IQ, OQ and PQ protocols which satisfy
FDA guidelines for qualification verification of equipment.
The Calibration Screen
To get to the calibration screen from the home page:
Press the
(Settings) button.
Be sure that all reference instruments are
calibrated to an appropriate standard.
Settings
Button