3.

CONNECTION TO PRESSURE SOURCES

The C9500 Series IS instruments are fi tted with

1

/

8

" BSP female connectors. The following checks

must be made before connecting the instrument to
a pressure source:-

Check the security of all fi ttings.

Check that the pressure source does not exceed the
ratings for the instrument as set out in Table 1.

Check that the gases to be measured will not damage
or corrode stainless steel, brass, glass, ceramic, silicon,
nickel, aluminium, gold or epoxy resin.

Note: Comark C9500 Series IS Pressure Measurement
Instruments are not suitable for absolute pressure
measurement.

Note: Pressures in excess of maximum over range
fi gure with respect to atmospheric pressure, should
not be applied to either port as this could damage the
instrument or represent a safety hazard.

TABLE 1

Instrument

Pressure Range

Maximum Over/

Under

Pressure

C9501/IS

0 to ±2 PSI/140mbar

6 PSI/400mbar

C9503/IS

0 to ±5 PSI/350mbar

15 PSI/1000mbar

C9505/IS

0 to ±30 PSI/2000mbar

90 PSI/6200mbar

C9507/IS

0 to ±100 PSI/6900mbar

200 PSI/13700mbar