Pid page – Despatch RFD1-42-2E with Protocol Plus User Manual

Page 40

Advertising
background image

32

PID Page


The PID Page contains parameters which control the response to the setpoint and
process variable input. To access the PID Page, enter the Setup Mode. Press the Page
key until PID is displayed. Press the Menu key. Each parameter can be changed by
pressing the Menu key until the desired parameter is displayed, and then pressing the

keys to change the value.

Menu Item

Display

Description

Range

Display units

PID Temp Unit

Set display units to °C or °F

°C or °F

Proportional
band

PID Prop Band

Set proportional band for tuning

1 to 56°C (1 to 100°F)

Integral reset

PID
Reset/Rep/Min

Set integral reset for tuning

0.0 to 100 seconds/repeat

Derivative
Rate

PID Rate In Sec

Set derivative rate for tuning

0.0 to 500 seconds

AutoTune

PID AutoTune

Enable auto tuning function

Disable, Enable



The AutoTune parameter disables or enables the AutoTune function. To utilize
AutoTuning:

1. Access the Setup Mode.

2. Press the Page key until the display reads AutoTune. Press the

key to enable

the AutoTune.

3. Press the Page key for three seconds to exit Setup Mode.

4. Cycle power to the instrument.

5. Set Manual Mode to run. The display will alternately display AutoTune and

Manual.


If the Manual Mode setpoint is less than 50 degrees higher than the actual process
temperature, the AutoTune function will create an error condition. This can be cleared
by either cooling off the process temperature or increasing the setpoint until there is
more than 50 degrees between them. Once the AutoTune function is allowed to
complete tuning, the AutoTune parameter will disable by itself.

If you wish to cancel the AutoTune function, press the STOP key, access the PID page
in Setup Mode, and set the AutoTune parameter to Disable.