Optional mrc5000 recorder setup – Despatch RFD1-42-2E with Protocol Plus User Manual

Page 75

Advertising
background image

67

Optional MRC5000 Recorder Setup


The temperature is retransmitted to the Recorder from the Controller. Set up the
Recorder as follows:

1.

Make sure that jumper JU1 is setup for the 5 VDC setting (see MRC Manual).


2.

Move the Mode switch to the PROG/TEST/CAL position, and Prog will be
displayed.


3.

Press the down arrow key twice and Inps will be displayed. Make sure the
settings are per the table below.


4.

Once all the settings have been changed, move the Mode switch to the RUN
position. The display on both the Recorder and controller should read the same.

Parameter Code

Degrees C

Degrees F

Inps

17

17

Icor

0

0

diSP

On

On

dPOS

0

0

EUU **

400

752

EUL **

0

32

ChUP

400

800 *

ChLO

0

0

DFF

1

1


* Note:

The 0 - 400 chart paper must be changed to the 0 - 800 chart paper.
Depending on the equipment, 0 - 600 paper can be used if the maximum
temperature is 500 degrees F.


** Note:

These values must mach the setting sRetOutLo and RetOutHi on the Control
page on the Protocol Plus controller (example: RetOutLo is 32, then EUL
must read 32).