Environmental Express Soil-Cell User Manual

Page 14

Advertising
background image

800.745.8218 / 843.576.1147 / www.environmentalexpress.com

Environmental Express

• 12

Soil-Cell:

Operation and Instruction Manual

Troubleshooting Your HotBlock

TM

The audible alarm has sounded
immediately after powering on
and the HotBlock

TM

will not heat.

There are two possible causes for your HotBlock

TM

to sound the alarm

immediately after the controller cycles through the self test. These are:
1. Your set point has been set to a value (≥) 15° less than ambient or current

set point temperature. Turn the set point to within 15° of the actual
temperature (blue numbers)

2. The controller is faulty. Call Environmental Express at 1-800-745-8218 for

more information.

The temperature has overshot
the set point and the audible
alarm has sounded.

The function of the fail-safe system is to cease heating of the HotBlock

TM

in the

event of a set point overshoot of 15°C and to alert the technician of the incident.
The HotBlock

TM

can be “fooled” into fail-safe mode if the set point is manually

changed to a value ≥15° below your current temperature. However, the primary
cause for the runaway temperature is a faulty relay that has exceeded its
useful life.
You may troubleshoot the relay by following these steps:

Caution: This procedure is a potential electrical hazard and should only
be performed by qualified personnel.

1. Power off your HotBlock

TM

.

2. Remove the bottom panel of the HotBlock

TM

by unscrewing the rubber feet.

3. Power the HotBlock

TM

on and allow it to overshoot your set point

temperature.

4. Locate the solid state relay mounted to the bottom panel.
5. Set your volt-meter to measure AC voltage.
6. Touch your red lead to terminal T1 (white wire) of the solid state relay and

touch your black lead to a ground source (e.g., the green/yellow wire from
the power module or an empty terminal on the terminal board).

7. If your volt-meter reads 110V-122V, then the relay is stuck in the “closed”

position and it must be replaced. (see parts list, page 11).

The temperature controller is
performing erratically or displays
an error message.
-OR-
The temperature controller is
flashing Er.L Attn
- Thermocouple
Fault

First reset default settings
1. Hold the up and down arrow buttons for six seconds until Ai Set appears
2. Hold the down arrow until glbl set is in the window
3. Press the green advance key to enter
4. Continue pressing the green advance key until none user appears.
5. Press the down arrow key until Set1 user appears.
6. Pressing the advance key will restore default settings.

To troubleshoot the thermocouple, follow these steps:
1. Power off your HotBlock

TM

.

2. Remove the bottom panel of your HotBlock

TM

by unscrewing the rubber feet.

3. Locate the thermocouple wires at terminals R1 (yellow) and S1 (red) of the

controller and remove using a small screwdriver.

4. Cut the exposed ends of the two wires.
5. Strip 1/4" of insulation from each wire and reconnect them to the

appropriate controller terminal and tighten.

6. Power on your HotBlock

TM

.

7. If your display continues to flash Er.1 Attn- the thermocouple is faulty and

must be replaced. (see parts list, page 11).

Advertising