Soil-Cell:

Operation and Instruction Manual

About Your HotBlock

TM

Information and Markings

Environmental Express HotBlocks

TM

provide an efficient method of digesting and storing water, wastewater, soil and sludge

samples for metals analysis. These innovative digestion systems allow samples to be digested in a corrosion-free environment.
In addition, samples are handled in a small area with minimal radiant heat loss. Users should be aware of potential dangers
from heating certain types of compounds. Such hazards may include explosion or the release of toxic or flammable gases.

Each HotBlock displays certain markings and symbols. All personnel working with the HotBlocks should have an

understanding of the following symbols and definitions:

• V = voltage
• ~ = alternating current
• Hz = frequency
• A = amperes

Unpacking Your HotBlock

TM

:

1. Remove the HotBlock

TM

from the shipping container by lifting from the bottom of the block. The lid should not be

used for lifting.

2. Your HotBlock

TM

is shipped with metal screws securing the bottom panel. The metal screws must be removed before

operating your HotBlock

TM

. Remove the metal screws and replace them with the PVC screws and rubber feet included

with your shipment. The corrosion-resistant PVC screws and rubber feet are designed to secure the bottom plate.

Important: Do not over tighten the PVC screws!

Installation Requirements

Locate the HotBlock

TM

under a fume hood with a minimum face velocity of 100fpm, and allow a minimum of 2" of space

on all sides. The following environmental conditions should be observed:

• Ambient temperature range:

5-30°C

• Ambient relative humidity:

0-90%RH

• Altitude:

sea level to 2500 meters

HotBlocks

TM

are rated as Pollution Degree 2 and Installation Category 2.

Electrical Requirements

Required Voltage: 120 volts, ~60Hz, 15A (all HotBlocks

TM

are also available in 240V with CE mark)

Power should not vary greater than±10%. Use the supplied heat-resistant power cord or equivalent to connect to the

power supply.

For safety reasons, a separate power receptacle should be provided for each unit in the system. Do not use extension

cords or outlet adaptors. Make certain that power outlets are earth-grounded at the grounding pin.
See individual specifications for each HotBlock model.

This symbol means Caution
Hot Surface
. The surface of the
HotBlock may be too hot to
safely touch with bare hands.

This symbol means Read
and become familiar with
instructions before operation of
instrument
.