4 technical data, Technical data – INFICON Ecotec E3000 Multigas-Sniffer-Leak Detector User Manual

Page 14

Advertising
background image

14

4 Description

Tr

an

sl

ation

o

f the

o

rig

in

al

operatin

g

in

str

u

ct

io

ns

Ecot

ec E3

000,

k

in

a

22en

1

-r

, 1

407

If the measuring value limit is exceeded, the display switches from green to red. You can
additionally set one speaker in the handle to output a signal and the LEDs in the handle to
flash or shine with maximum brightness,

see “6.3 Settings before measurements”, page 28

.

4.4

Technical data

Table 2: Technical data

Mechanical data

Dimensions (W

 H  D)

610 mm

 370 mm  265 mm

Weight

34 kg

Ambient conditions

Permissible ambient temperature (during operation)

10 °C to 45 °C

Permissible storage temperature

-20 °C to 60 °C

Max. relative humidity up to 31 °C

80 %

Max. relative humidity from 31 °C to 40 °C

Linearly dropping from 80 % to 50 %

Max. relative humidity above 40 °C

50 %

Pollution degree

II
(according to IEC 61010/part 1: “Usually, only non-
conducting soiling may occur. However, temporary
conductivity caused by condensation is tolerable at times.“)

Max. altitude above sea level

2000 m

Electrical data

Supply voltages and frequencies

Country-specific, see rating plate on device rear

Power consumption

 300 VA

Protection class

IP 20

Overvoltage category

II

Mains fuse

2

 4 A slow-blow

Power supply cable

2.5 m

Noise level

< 54 dBA

Physical data

Minimum detectable leak rate

R134a

0.05 g/a (0.002 oz/yr)

R600a

0.05 g/a (0.002 oz/yr)

Helium

< 1

 10

-6

mbar l/s

Measurement range

6 decades

Detectable masses

2 to 200 amu

Mass spectrometer

Quadrupol mass spectrometer

Ion source

2 cathodes

Time constant of the leak rate signal

< 1 s

Gas flow through the capillary
Measured at 1 atm (1013 mbar) at sea level. The flow rate
changes with geographical height and barometric pressure.

120 to 200 sccm

Time until ready for operation

< 2 min

Advertising