WARNING - Risk Of Electric Shock
Do not use if the "L" connector is broken.
WARNING - Risk Of Electric Shock
Connect the Hot Swap Cable "L" connector to the battery
first. Then, plug the power connector into the HAPSITE
24 V(dc) power port.
2
Plug the power connector into the 24 V(dc) power port on the HAPSITE. The
depleted internal battery can now be replaced. After replacing the internal
battery, disconnect the Hot Swap Cable. See
.
CAUTION
Do not open the HAPSITE front panel in a contaminated
or wet environment.
Figure 2-75 Power Connector
HAPSITE
24 V(dc)
Power