6.4 Replacing the Crystal

The procedure for replacing the crystal is basically the same with all transducers,
except the Cool Drawer, RSH-600, CrystalSix and Crystal12.

CAUTION

Always use clean nylon lab gloves and plastic tweezers
for handling the crystal (to avoid contamination which
may lead to poor adhesion of the film to the electrode).

Do not rotate the ceramic retainer assembly after it is
seated (as this will scratch the crystal electrode and
cause poor contact).

Do not use excessive force when handling the ceramic
retainer assembly since breakage may occur.

NOTE: Certain materials, especially dielectrics, may not adhere strongly to the

crystal surface and may cause erratic readings.

NOTE: Thick deposits of some materials, such as SiO, Si, and Ni will normally

peel off the crystal when it is exposed to air, as a result of changes in
film stress caused by gas absorption. When you observe peeling, replace
the crystals.

6.4.1 Front Load Single Sensors

Follow the procedure below to replace the crystal in the Front Load Single sensor
(see

Figure 6-3

):

1

Gripping the crystal holder with your fingers, pull it straight out of
the sensor body.

2

Gently pry the crystal retainer from the holder (or use the Crystal Snatcher;
see

Figure 6-8 on page 6-23

).

3

Turn the retainer over and the crystal will drop out.

4

Install a new crystal, with the patterned electrode face up.

5

Push the retainer back into the holder and replace the holder in
the sensor body.