INFICON SKY CDG045D 0.05 ... 1000Torr / mbar User Manual

Page 24

Advertising
background image

24

tina51e1-c (2011-12)

Operate the gauge for at least ¼ hour (until the signal is

stable).

Briefly press the <ZERO> button with a pin (max.

ø1.1 mm). The zero adjustment runs automatically. The
<STATUS> LED blinks until the adjustment (duration ≤8 s)
is completed.

Press the button briefly

max. ø1.1 mm

After zero adjustment, the gauge automatically returns

to the measurement mode.

The <STATUS> LED blinks green if

• the signal output is negative (< -20 mV) when the final pres-

sure has been attained

• the zero adjustment has failed.

4.2.2

<ZERO> Adjustment with Ramp Function

The ramp function allows to adjust the zero at a known reference
pressure within the measurement range of the gauge.

Advertising