tina51e1-c (2011-12)

4.3.1

Adjusting the Setpoints

The setpoints can be adjusted via

• the buttons on the gauge,
• the diagnostic port (→  [4]),
• the RS232C interface (→  [4]).

DANGER

DANGER: malfunction
If processes are controlled via the signal output,
keep in mind that by pushing the <SP> button the
measurement signal is suppressed and the cor-
responding threshold value is output instead. This
can cause malfunctions.
Push the <SP> button only if you are sure that no
damages can arise from a malfunction.

Adjusting Setpoint <1>

Push the <SP> button with a pin (max. ø1.1 mm). The

gauge changes to the switching function mode and outputs
the current threshold value at the measurement value
output for about 10 s (LED <1> blinks).