4.2.1

<ZERO> Adjustment

Evacuate the gauge to a pressure according to the table

below:

F.S.

Recommended final pressure for

zero adjustment

1100 mbar
1000 Torr/mbar
200 Torr/mbar
100 Torr/mbar
20 Torr/mbar
10 Torr/mbar

2

Torr/mbar

1

Torr/mbar

0.25 Torr/mbar
0.1 Torr/mbar


<5

×10

-2

Torr

<10

-2

Torr

<5

×10

-3

Torr

<10

-3

Torr

<5

×10

-4

Torr

<10

-4

Torr

<5

×10

-5

Torr

<10

-5

Torr

<5

×10

-6

Torr

<6.65

×10

0

Pa

<6.65

×10

0

Pa

<1.33

×10

-0

Pa

<6.65

×10

-1

Pa

<1.33

×10

-1

Pa

<6.65

×10

-2

Pa

<1.33

×10

-2

Pa

<6.65

×10

-3

Pa

<1.33

×10

-3

Pa

<6.65

×10

-4

Pa

<5

×10

-2

mbar

<5

×10

-2

mbar

<10

-2

mbar

<5

×10

-3

mbar

<10

-3

mbar

<5

×10

-4

mbar

<10

-4

mbar

<5

×10

-5

mbar

<10

-5

mbar

<5

×10

-6

mbar

If the final pressure in the gauge is too high for zero ad-
justment (>25% of the F.S.), the zero cannot be reached
and the <RUN> LED flashes. If this is the case, activate
the factory setting and adjust the zero again (

→ 27).

Operate the gauge for at least ¼ hour (until the signal is

stable).