tina49e1-c (2009-06)

Briefly press the <ZERO> button with a pin (max.

ø1.1 mm). The zero adjustment runs automatically. The
<STATUS> LED flashes until the adjustment (duration
≤8 s) is completed.

Press the button briefly

max. ø1.1 mm

After zero adjustment the gauge automatically returns to

measurement mode. The <RUN> LED lits.

The zero can also be adjusted via the RS232C interface

(

→ [4]).

The <RUN> LED flashes if

• the signal output is negative (< -20 mV) when the final pres-

sure has been attained

• the zero adjustment has failed.

4.2.2

<ZERO> Adjustment with Ramp Function

The ramp function allows to adjust the zero at a known reference
pressure within the measurement range of the gauge.