29

OverTemp 87.0°C

SubTemp 83.0°C

Setpoint3 85.0°C

The warning functions are only triggered when the actual bath temperature,
after start from the „OFF“ or „rOFF“ mode, lies within the set limits for 3
seconds.


Setting range:
20 °C ... 220 °C

SafeTemperature: Indicated is the cut-out value set with a screwdriver on the

excess temperature protection device.

This safety installation is independent of the control circuit. When the
temperature of the bath fluid has reached the safety temperature, a complete
shutdown of the heater and pump is effected.
The alarm is indicated by optical and audible signals (continuous tone).
The following error message appears on the VFD COMFORT-DISPLAY:

XXXXX

ALARM
CODE 14

Press

for help on the LCD

Working temperature
exceeds the adjusted
safety temperature.

SafeSens: Indicated is the temperature value of the safety sensor.