Troubleshooting guide / Error messages

64

ALARM

CODE 06

Difference between
working- and safety-
temp. Bigger than
permissible value.

 Defect of the working or excess temperature sensor.

Working temperature and excess temperature sensors report a
temperature difference of more than 35 K.

ALARM

CODE 07

The internal I2C
bus-system shows
a failure.


 Other errors (I

2

C-BUS errors)

ALARM

CODE 12

Error in A/D-
converter of the
internal measuring
system.

 Error in A/D converter

ALARM

CODE 14

Working temperature
exceeds the adjusted
safety temperature.

 Excess temperature sensors defect.
 The excess temperature value lies below the working temperature

setpoint. Set the excess temperature to a higher value.

ALARM

CODE 15

External control
selected, but ext.
sensor not connected
or defective.

 External control selected, but external Pt100 sensor not connected

or defect.

WARNING

CODE 20

Cooling is affected
or ambient temperature
to high
Clean condenser

 Cooling of the condenser is affected. Clean air-cooled condenser.

Check the flow rate and cooling water temperature on water-cooled
condenser.

WARNING

CODE 21

Compressor does not
work. Compressor
will be automatic-
ally reconnected.

 Compressor does not work.

After a short cooling interval, the compressor motor will be
automatically reconnected and the message "CODE 21" no longer
appears.

ALARM

CODE 33

Internal safety-
temperature sensor
short-circuited or
interrupted.

 Cable of the excess temperature sensor interrupted or short-circuited