Operating procedures – JULABO FLW11006 Recirculating Coolers User Manual

Page 20

Advertising
background image

20

7. Operating

procedures

7.1. Bath fluids

Caution:

No liability for use of other bath liquids!
Please contact JULABO before using other than recommended bath fluids. JULABO
takes no responsibility for damages caused by the selection of an unsuitable bath fluid
Do not use alcohols.

Water:

The quality of water depends on local conditions.
 Due to the high concentration of lime, hard water is not suitable for temperature

control because it leads to calcification in the bath.

 Ferrous water can cause corrosion - even on stainless steel.
 Chloric water can cause pitting corrosion.
 Distilled and deionized water is unsuitable. Their special properties cause corrosion

in the bath, even in stainless steel.

No liablity for use with water. Danger of freezing at working temperatures <5 °C.

Recommended bath fluids:

Bath fluids

Temperature range

soft/decalcified water

5 °C to 80 °C

See website for list of recommended bath fluids.
Contact: www.JULABO.de




7.2. Power

connection

Caution:

 Only connect the unit to a power socket with earthing contact (PE – protective earth)!
 The power supply plug serves as safe disconnecting device from the line and must

be always easily accessible.

 Never operate equipment with damaged mains power cables.

 Regularly check the mains power cables for material defects (e.g. for cracks).
We disclaim all liability for damage caused by incorrect line voltages!

Make sure that the line voltage and frequency match the supply voltage specified on the
type plate.

Advertising