Electrical connections – JULABO FLW11006 Recirculating Coolers User Manual

Page 26

Advertising
background image

26

10. Electrical

connections

Notice:

Use shielded cables only.

The shield of the connecting cable is electrically connected to the plug housing.

The unit ensures safe operation if connecting cables with a maximum length of 3 m are
used. The use of longer cables does not affect proper performance of the unit, however
external interferences may have a negative impact on safe operation.


RS232 serial interface
This port can be used to connect a computer with an RS232 cable for remote
control of the recirculating cooler.

9

6

5

1

RS232C

Pin assignments RS232:

Pin

2

RxD

Receive

Data

Pin

3

TxD

Transmit

Data

Pin 5

0 V

Signal GND

Pin 7

RTS

Request to send

Pin 8

CTS

Clear to send

Pin 1; 4; 6, 9 Reserved - do not use!

Accessories:

Order No.

Description

8 980 073

RS232 interface cable 9-pol./9-pol. , 2.5 m

8 900 110

USB interface adapter cable


Alarm output

Potential-free change-over contact for external alarm signal.

Pin 2 and 3 are connected in case of an alarm.
Pin 2 and 1 are connected in normal condition or mains switch "Off".

or


Switching capacity

max. 30 W / 30 VA

Switching voltage

max. 30 V

/

Switching current

max. 1 A

Advertising