Counter-cooling – JULABO SL-26 Heating Circulators User Manual

Page 23

Advertising
background image

HL / SL

23

5.5. Counter-cooling

Caution:

Securely attach all tubing to prevent slipping.

Observe the laws and regulations of the water distribution company valid in the
location where the unit is operated.

T16A

T16A

T1.25A

ext

Pt100

SERIAL

ALARM

REG+
E-PROG

STAND-BY

20

For applications near the ambient temperature, the cooling coil (20)
must be connected to the water mains.

 Using tubing, connect the cooling coil to the tap water supply,

and lead the tap water in a sink through the return connector.

A specific water flow rate of 750 ml/minute (15 °C) is sufficient
to compensate for the characteristic temperature.

For applications near ambient temperature (20 °C), the cooling
water temperature should at least be 5 °C below the working
temperature.

T16A

T16A

T1.25A

ext

Pt100

SERIAL

ALARM

REG+
E-PROG

STAND-BY

V

Controlled counter-cooling

A) Accessory: Solenoid valve for cooling water

Order No. 8 980 703

Use a solenoid valve (V) for controlled counter-
cooling to reduce tap water consumption. The
necessary cooling pulse is emitted via the socket
(14).

In menu >Inputs/Outputs< set the menu item >SV-
Out< to >C-Pulse<.

see page 65

In menu >Limits< set the menu item
>CoolingMax< to >100%<.

see page 62

FD200 Flow-through cooler

Order No. 9 655 825


The FD200 is suitable as cooling device for liquids in closed circuits (loops).
This unit is integrated in the return flow circuit from the external system
connected to a circulator, and continuously withdraws heat from the liquid.