Installation of the magnetic valve
JULABO Order-No. 8 980 704 3

Magnetic valve

• Fix the expansion tank (43) to the connection (41) (open

end wrench SW19)

• Screw the magnetic valve to the entry for cooling water

(45). Respect the flow direction (arrow)
Connect the cooling water to the magnetic valve and to the
outflow of cooling water (44) by means of tubes 8 mm ID.

• Fix the control cable (56) at the control exit (56a) and srew

tightly.


50

51

Adapter to reduce the pump pressure

JULABO Order-No. 8 970 802

With this adapter the pump pressure at the pump connection
(51) can be reduced from 1,2 bar to
0,8 bar

–0,1

.

Assembly
• Fix the adapter to reduce the pump pressure to the pump

connections (50, 51) (open end wrench SW19).

Notice:

When using pressure sensitive glass vessels, it is absolutely necessary to work with

an adapter to reduce the pump pressure.


Caution:

Securely attach all tubing to prevent slipping.