12.4. Status messages

Status messages

Description

00 MANUAL STOP

Recirculating cooler in „OFF“ state.

01 MANUAL START

Recirculating cooler in keypad control mode.

02 REMOTE STOP

Recirculating cooler in „r OFF“ state.

03 REMOTE START

Recirculating cooler in remote control mode.

12.5. Error messages

Error messages

Description

-01 LOW LEVEL ALARM

Low liquid level alarm or
Excess temperature protector alarm

-03 EXCESS TEMPERATURE

WARNING

High temperature warning.

-04 LOW TEMPERATURE WARNING

Low temperature warning.

-05 WORKING SENSOR ALARM

Working temperature sensor short-circuited or interrupted.

-06 SENSOR DIFFERENCE ALARM

Sensor difference alarm.
Working temperature and safety sensors report a
temperature difference of more than 35 K.

-07 I

2

C-BUS ERROR

Internal error when reading or writing the I

2

C bus.

-08 INVALID COMMAND

Invalid command.

-09 COMMAND NOT ALLOWED IN

CURRENT OPERATING MODE

Invalid command in current operating mode.

-10 VALUE TOO SMALL

Entered value too small.

-11 VALUE TOO LARGE

Entered value too large.

-12 TEMPERATURE MEASUREMENT

ALARM

Error in A/D converter.

-13 WARNING : VALUE EXCEEDS

TEMPERATURE LIMITS

Value lies outside the adjusted range for the high and low
temperature warning limits. But value is stored.

-14 EXCESS TEMPERATURE

PROTECTOR ALARM

Excess temperature protector alarm

-15 EXTERNAL SENSOR ALARM

External control selected, but external Pt100 sensor not
connected.

-33 SAFETY SENSOR ALARM

Excess temperature sensor short-circuited or interrupted.

-38 EXTERNAL SENSOR SETPOINT

PROGRAMMING ALARM

Ext. Pt100 sensor input without signal and setpoint
programming set to external Pt100.

-40 NIVEAU LEVEL WARNUNG

Low liquid level warning in the internal reservoir.