Abnormality & Troubleshooting

Abnormality

Reason

Action

The cuff does not inflate or the display
shows low pressure during inflation

LCD screen shows battery symbol

LCD screen shows abnormal result

LCD shows “EE”

No response when buttons pressed

Pressure loss due to air leakage in

tube or cuff

Rubber tube is not connected

properly

Batteries are low

Cuff is not tight or is

positioned incorrectly
Subject was excited/nervous or

spoke/moved during measurement

Wrong position

Some interference during inflation

or incorrect operation during

measurement
Insufficient inflation

Some interference during inflation

or incorrect operation during

measurement

Incorrect operation

Check the connection between the tube

and monitor

Reconnect the rubber tube properly

Change the batteries - refer to the ‘Loading

the Batteries’ section

Tighten the cuff and refer to the section

‘Setting up your Blood Pressure Monitor’
Re-test. Do not speak and take 2-3 deep

breaths before to relax. Stay calm and

remain steady.
Adjust your position - refer to part a on

page 7

Refer to the section ‘Using your Blood

Pressure Monitor’

Re-measure

Re-measure

Take out the batteries for a few minutes,

then re-insert them, following directions in

section ‘Loading the Batteries’.