SERVICE AND TROUBLE-SHOOTING

Section

9

TEMPERATURE

Temp-1 Temperature too high-

display and reference
thermometer don’t
match

1/ Controller failed on – call Customer Service.
2/ Wiring error – call Customer Service.

Temp-2 Display reads "HI" or

"400"+

Probe is unplugged or broken – Call Customer Service

Temp-3 Chamber

temperature

spikes over set point
and then settles to set
point

Recalibrate – see section 7.4

Temp-4 Temperature too low-

display and reference
thermometer don’t
match

13