Mechanical – Labnet 611D Laboratory Incubators User Manual

Page 15

Advertising
background image

15

including motors, arcing relays or radio transmitters.
7/ Bad connection on temperature sensor or faulty
sensor – check connectors for continuity and
mechanical soundness while watching display for
erratic behavior; check sensor and wiring for
mechanical damage.
8/ Bad connections– check connectors for mechanical
soundness and look for corrosion around terminals or
signs of arcing or other visible deterioration.

Temp-9 Will not maintain set

point

1/ Assure that set point is at least 5 degrees over
ambient.
2/ See if ambient is fluctuating – check for adjacent
open doors or HVAC duct openings, stabilize ambient
conditions.

Temp-10 Display and reference

thermometer don’t
match

1/ Calibration error – see section 7.4.
2/ Temperature sensor failure – evaluate if pilot light is
operating correctly.
3/ Controller failure – evaluate if pilot light is operating
correctly.
4/ Allow at least 24 hours to stabilize at set point
temperature.
5/ Verify that reference thermometer is certified.

Temp-11 Can't adjust set points

or calibration

1/ Turn entire unit off and on to reset.
2/ If repeatedly happens, call Customer Service.

Temp-12 Calibrated at one

temperature, but not at
another

This can be a normal condition when operating
temperature varies widely. For maximum accuracy,
calibration should be done at or as close to the set
point temperature as possible.



MECHANICAL

Mech-1 Door not sealing

1/ Check physical condition of gasket.
2/ Tighten door latch till it pulls glass in.
3/ Assure that gasket is in original location.
4/ Adjust hinge blocks or twist the door.
5/ Confirm that unit has not been damaged and body
is square.

Mech-2 Motor doesn't move