Leica ASP6025

Changing reagent names or reagent groups

• Highlight the reagent the name or group of which you wish to change.
• Press the corresponding button symbol in the headline.
• In the entry window (or via the keyboard), enter the new reagent group allocation / the new

reagent name.

• Press OK to save the new reagent group / reagent name.

Entering / modifying reagent thresholds
If warning thresholds are required for a certain reagent, enter them as follows:
• Highlight the reagent to be modified, either by pressing the reagent name or using the

UP/DOWN buttons.

• In the header line, touch the parameter to be changed - the corresponding entry screen appears

(

Fig. 132

).

• Enter a new threshold value or - if no warning is desired - press CLEAR to remove the threshold

altogether.

• Press OK to confirm.

Changeable parameters

Fig. 132

Fig. 133

Configuring the reagent list

(continued)

5. Operation

Allocation of reagents to the correct reagent group is the basis for compatibility monitoring. Allocation to the

wrong reagent group can lead to reagent cross contamination.

The respective threshold value applies for all reagent stations containing the

same reagent.

For additional information, refer to Chapter 5.2.2

If a reagent is already used in a program, it can neither be renamed nor can it

be allocated to another reagent group!

The corresponding symbols will be disabled (i.e. they will not be surrounded by

a blue border).

If a reagent is renamed, its allocation to a station will be deleted.