Leica Biosystems Bond Oracle HER2 IHC System User Manual

Page 15

Advertising
background image

Page 15 of 23

English

Leica Biosystems Bond Oracle HER2 IHC System Instructions for Use TA9145 EN-CE-Rev_E 18/06/2013

English

2x2 Concordance Results

In this primary analysis the test results from the two tests (Bond Oracle HER2 IHC System

and Dako HercepTest) are categorized as negative (0,1+) or positive (2+, 3+). The frequencies

of four possible combinations are displayed in a 2x2 table format (see Table 6). Then, the

overall concordance rate based on this 2x2 table was calculated accompanied by a 95% exact

confidence interval (based on the binomial distribution).
The null hypothesis (H

0

), which the success criteria are set against, is that concordance is no

greater than 75%.
The observed agreement for 431 samples between the two tests in a 2x2 analysis show a

concordance of 92.34% (398/431) with a 95% CI of 89.42% - 94.67%. This data supports rejection

of the null hypothesis (H

0

) that agreement was no greater than 75% with a p-value<0.0001.

The percentage Positive Agreement (sensitivity) or the ability of Bond Oracle HER2 IHC System

to correctly identify HercepTest positive cases (the percentage of specimens scored positive

by both Bond Oracle HER2 IHC System and HercepTest out of all the HercepTest positive

cases) was 84.87% (129/152) with a 95% CI of 78.17%-90.16%. The percentage Negative

Agreement (specificity) or the ability of the test to correctly identify HercepTest negative cases

(the percentage of specimens scored negative by both Bond Oracle HER2 IHC System and

HercepTest out of all the HercepTest negative cases) was 96.42% (269/279) with a 95% CI of

93.51%-98.27%.

HercepTest

Negative

Positive

Totals

Bond Oracle HER2
IHC System

Negative

269

23

292

Positive

10

129

139

Totals

279

152

431

2x2 Concordance (95% CI) = 92.34% (89.42 to 94.67%); p<0.0001

Table 6. 2x2 concordance of Bond Oracle HER2 IHC System with HercepTest

3x3 Concordance Results

Data was grouped as negative (0 or 1+), equivocal (2+) or positive (3+) for 3x3 analysis and

showed a concordance of 86.54% (373/431) with a 95% CI of 82.95% to 89.62 %. Therefore, the null

hypothesis (H

0

) that agreement was no greater than 75% was rejected with a p-value<0.0001.

The percentage Positive Agreement for 3+ (the percentage of specimens scored 3+ positive

by both Bond Oracle HER2 IHC System and HercepTest out of all the 3+ HercepTest positive

cases) in this study was 73.33% (66/90) with a 95% CI of 62.97% to 82.11%. The percentage

Negative Agreement was 96.42% (269/279) with a 95% CI of 93.51% to 98.27. See Table 7.

HercepTest

Negative
(0 or 1+)

2+

3+

Totals

Bond Oracle
HER2 IHC
System

Negative

(0 or 1+)

269

23

0

292

2+

10

38

24

72

3+

0

1

66

67

Totals

279

62

90

431

3x3 Concordance (95% CI) = 86.54% (82.95% to 89.62 %); p<0.0001

Table 7. 3x3 concordance of Bond Oracle HER2 IHC System with HercepTest