2 ct0 & ct1 commands – Loadstar Sensors DQ-4000 User Manual

Page 16

Advertising
background image

DQ-4000

DQ-4000 User's Guide

Page: 16 / 21

P/N : 033-02061 Rev. 1

4.2 CT0 & CT1 Commands

Function:

Tares (zero) the selected load cell

Summary:

The load on the selected load cell is set as the zero values and all subsequent
load readings are relative to this gross zero value. CT0 is a “hard” tare that
saves the values to the flash memory. While CT1 is a “soft” tare that the value is
not preserved when the power is lost. The commands are issued without any
other parameters.

Example:

CT0 (tares currently selected sensors and save the value to the flash memory)

CT1 (tares currently selected sensors, but without saving the values to the flash
memory)

4.3 O0S0 & O0S1 Commands

Function:

Output the raw data (frequencies and temperature) for the currently
selected channel

Summary:

These commands output the raw data read from the selected sensor. The data
is output in 6 columns: Compensated frequency, Sensor frequency, Reference
frequency, Temperature (ºC x 100), millisecond timer, and Computed load
(millipounds).

O0S0 outputs the data continuously, while O0S1 outputs it once. The streaming
data generated by an O0S0 can be stopped by pressing [Enter].

The O0S0 and O0S1 are valid when individual sensor is selected, using S1, S2,
S3 or S4 command. It is not valid when all sensors are selected using S0
command.

Example:

O0S0 (Starts streaming raw data from current channel)