GROUNDING METHOD AND ELECTRICAL REQUIREMENTS

• Do not splash or spray water from a hose on the refrigera-

tor! Doing so may cause an electrical shock, which may

result in severe injury or death.

• This unit should not, under any circumstances, be installed

to an un-grounded electrical supply.

Grounding Method

This product is factory equipped with a power supply cord that

has a three-pronged, grounded plug. It must be plugged into a

mating grounding type receptacle in accordance with the Na-

tional Electrical Code and applicable local codes and ordinances

(see Figure 5). If the circuit does not have a grounding type re-

ceptacle, it is the responsibility and obligation of the customer to

provide the proper power supply. The third ground prong should

not, under any circumstances, be cut or removed.

Electrical Requirements

Electrical Requirements: A grounded 115 volt, 15 amp dedicated

circuit is required.

The unit must be installed according to your local building codes

and ordinances.

Do not use an extension cord with

this appliance.