D24034 | Date 10/2014

Delivery, transport and setting up

3.6 Setting up

Warning!

Due to its centre of gravity, the appliance can fall over to the front

and injure you or other people. Always attach the appliance to a wall

with the tilt protection (see page 20). If this cannot be done due

to space problems, do not operate the appliance and do not open

the door. Contact the Memmert service team (see page 2).

3.6.1 Prerequisites

The installation site must be flat and horizontal and must be able to reliably bear the weight

of the appliance (see Technical data on page 13). Do not place the appliance on a flamma-

ble surface.
Depending on the model (see nameplate), a 230 V or 115 V power connection must be avail-

able at the installation site.
The distance between the wall and the rear of the appliance must be at least 15 cm. The clear-

ance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby

appliances must not be less than 5 cm (Fig. 7). Sufficient air circulation in the vicinity of the

appliance must be guaranteed at all times.