3 loading the appliance – Memmert IPS Peltier-cooled incubators User Manual

Page 23

Advertising
background image

D24034 | Date 10/2014

23

Operation and control

5.3 Loading the appliance

Warning!

When loading the appliance with an unsuitable load, poisonous or

explosive vapours or gases may be produced. This could cause the

appliance to explode, and people could be severely injured or poi-

soned. The appliance may only be loaded with materials which do

not form any toxic or explosive vapours when heated up and cannot

ignite (see also Intended use on page 8). If there is any doubt as

to the composition of materials, they must not be loaded into the

appliance.

Caution:

Check the chamber load for chemical compatibility with the materials of the appliance

(see page 11).

Insert the sliding steel grids or sliding shelves. The maximum number or grids / shelves and the

load capacity are specified in the technical data overview from page 13 .
The chamber must not be

loaded too tightly, so that

proper air circulation in the

interior is guaranteed. Do

not place any of the chamber

load on the floor, touching

the side walls or right below

the ceiling of the chamber

(Fig. 11, see also the "correct

loading" sticker on the appli-

ance).
In case of improper loading

(chamber loaded too tightly),

the set temperature may

be exceeded or it may take

longer until it is reached.

To achieve the correct heating capacity, the type of slide-in unit used – Grid or Shelf –

must be set in the menu under

SETUP

(see page 34).

Loading – Models 30 - 1060

Open / close the door

E24899

Memmert GmbH + Co. KG | P.O. Box 1720 | D-91107 Schwabach | Tel. +49 (0) 9122 / 925 - 0 | Fax +49 (0) 9122 / 145 85 | E-Mail: [email protected]

www.memmert.com | www.atmosafe.net

Fig. 11

Correct placement of the chamber load