Series 248 Actuator Product Information

40

How to Precharge the Static Support

Operation

Charger Connection

Rapid flow rates with pressure differentials of more than 2.1 MPa (300 psi)
across the input check valve can occur when you open the regulator shut-
off valve.

Damage to the valve seals can result from rapid flow rates.

Open the regulator shut-off valve only far enough to permit a gradual transfer of
gas.

Extension Hose

r

Bleed Valve

Input

Check Valve

Open

Close

Nitrogen Supply Hose

Regulator Output

Pressure Gage

THESE ITEMS SUPPLIED BY USER

Nitrogen Bottle

Pressure Gage

Regulator

Shut-Off Valve

Regulator Output

Pressure Valve

Nitrogen

Bottle Valve

Chuck Valve

Gage Protector

(factory set to limit

pressure to the gage

to approximately

1.4 MPa (200 psi)

High Pressure Gage

0-21 MP (0-3000 psi )

Poppet Type Valve

To Open and Close

Use Locknut

Open

Valve
Stem

Nitrogen

Bottle

Low Pressure Gage

0-2.1 MPa (0-300 psi)

Locknut

Close

Open

Close

Open

Close

Open

Close

CAUTION